If purchased, un-wrap the mechanical stage
(MA380/05). Locate the two holes on the bottom side
of the plain stage on the right hand side. The two
thumbscrews will attach the mechanical stage to the
plain stage as shown in Figure 6. Tighten the
thumbscrews snugly. Verify the smooth operation of
the mechanical stage by turning the coaxial controls
and checking for full un-restricted travel and
movement.
Plug the power cord into the microscope as shown in
Figure 7 and the other end into a grounded outlet. Your
TC has a voltage sensing power supply so it can use
voltage from any country in the world from 100 volts to
240 volts AC.
The mains power cord should only be plugged into a known grounded outlet.
Contact your facilities technician if you are unsure of your mains outlet status. A simple
outlet tester can be used to verify correct outlet polarity and the presence of a grounded
circuit.
If no other accessories are going to be installed, the instrument is now ready for use.