Possible Incident Light Fluorescence Operational Problems
If normal adjustments are not getting the results you expect, check these:
• Weak fluorescence or image intensity: specimens improperly stored or too old
• Rapid fading of specimens i.e.: photobleached
• Unspecific or erroneous filter combination
• Numerical aperture of objectives too low
• Eyepiece magnification too high
• Failing or old mercury lamp
• Ambient light or room too bright
• Secondary light due to reflection at condenser
• Low contrast image due to excitation bandwidth too wide
• Unspecified staining
• Fluorescing mounting medium
• Auto-fluorescence of objective or immersion oil
• Dirty optics
TC-5500 and TC-5600 Epi-Fluorescence Objectives
Epi-Fluorescent Objectives - Planachromat - Infinity Corrected - F = 200mm
MA849
TC Planachromat F4X objective ∞/1.0, NA: 0.13, WD = 17.5mm, F.L.= 50.0 (optional)
MA850
TC Planachromat F10X objective ∞/1.0, NA: 0.30, WD = 7.5mm, F.L.= 20.0 (included)
MA851
TC Planachromat F20X objective ∞/1.0, NA: 0.40, WD = 7.5mm, F.L.= 10.0 (included)
MA852
TC Planachromat F40X objective ∞/1.0, NA: 0.60, WD = 2.9mm, F.L.= 5.0 (optional)