5.4 Fluorescence Lamp does not work
LAMP HOUSE MAY BE HOT TO THE TOUCH.
Do not touch the glass envelope of the lamp at any time. Keep the protective sleeve or
bag of the lamp during installation and remove it right after installation.
• Make sure that all cable connections between the lamp house, power unit and
mains are all properly established.
• Check if the lamp house power unit main fuse is intact.
• Check the logged hours of the lamp for an excess of 400 hours.
5.5 Replacing the Fluorescence Lamp
Refer to the companion manual of the Nobska Lamphouse for more on
troubleshooting and replacement of the 100W Mercury Lamp.
Replacing the mercury HBO lamp within the Nobska housing is not very complex but
there is a followed procedure that will make the process easier and less dangerous.
If the glass envelope of the mercury bulb happens to break open, you and the immediate
environment could be exposed to mercury which is a known biohazard.
Please refer to the manual which came with your Nobska Lamp House for information
and instructions for replacing, adjusting and burning in your new mercury lamp.